Compact, quiet-running, autonomous device designed for artificial lung ventilation with an oxygen-air mixture and oxygen inhalation in ambulances, directly on the scene and in the field, as well as in clinical departments of surgery, intensive care, and intensive care wards.
The device performs the following functional actions:
a) controlled artificial respiration with adjustable respiration rate and minute ventilation;
b) oxygen inhalation with actuation from the patient's respiratory effort “on call” and a constant supply of oxygen.
c) short-term auxiliary ventilation of the lungs with oxygen “on call”, i.e. under the influence of the patient's respiratory effort;
g) the ability to work complete with the apparatus of inhalation anesthesia;
d) the possibility of attaching a gas mask for mechanical ventilation in poisoned places.
The device works from the built-in two-liter cylinder, as well as from an external source of oxygen and compressed air.
It is possible to charge the cylinder without removing it from the device.
It is easy to operate and maintain, provides disassembly of the respiratory circuit for disinfection and sterilization, and is resistant to mechanical and climatic influences.
Technical specifications
Parameter name |
Description |
Supply pressure, MPa (kgf/cm2): - from the oxygen tank |
19,6 |
- from an external compressed source oxygen or air |
from 0.3 to 0.6 (from 3 to 6) |
Minute ventilation, l / min |
10-25 |
Respiratory rate min -1 |
10 - 60 |
The ratio of the duration of inhalation and exhalation |
min 1:1,3 max 1: 2 |
Maximum safe pressure in the respiratory circuit, kPa (mm w.g.) |
5 (500) |
The gas flow in the ventilation mode, l /min, max: |
|
- 10 l / min - 25 l / min |
4 10 |
Oxygen concentration in the mixture, % |
|
- in ventilation mode - in the mode of inhalation |
35 (10-5) l (min) |